General specifications
Temperature
- Operating: 10 °C to 40 °C (50 °F to +104 °F)
- Storage: -25 °C to +50 °C (-13 °F to +122 °F)
Humidity
10 % to 80 % non-condensing
Altitude
2,000 meters (6,562 ft)
Dimensions (LxWxH)
14.0 cm x 20.6 cm x 4.5 cm (5.5 in x 8.2 in x 1.8 in)
Display
LCD greyscale display
Communication
USB device upstream port
Power
Two 9 V alkaline batteries
Battery life
8 hours continuous operation
Weight
0.47 kg (1 lb, 4 oz)
Safety standards
IEC 61010-1 3rd Edition; Pollution degree 2
Certifications
CE, CSA, C-TICK N 10140, RoHS
Electromagnetic compatibility (EMC)
IEC 61326-1:2012
Detailed specifications
Normal-sinus-rhythm waveforms
ECG reference
The ECG amplitudes specified are for Lead II (calibration), from the baseline to the peak of the R wave. All other leads are proportional.
Normal sinus rhythm
12-lead configuration with independent outputs referenced to right leg (RL). Output to 10 universal ECG Jacks, color-coded to AHA and IEC Standards
Amplitude
0.05 mV to 0.45 mV (0.05 mV steps); 0.5 mV to 5.5 mV (0.5 mV steps)
Amplitude accuracy
± 2 % of setting Lead II. All other leads ± 5 %
ECG rate
30, 40, 45, 60, 80, 90, 100, 120, 140, 160, 180, 200, 220, 240, 260, 280 and 300 BPM
Rate accuracy
± 1 % of setting
ECG waveform selection
Adult (80 ms) or pediatric (40 ms) QRS duration
ST-segment elevation
- Adult mode only: -0.8 mV to +0.8 mV (0.1 mV steps)
- Additional steps: +0.05 mV and -0.05 mV
Power-on default
80 BPM, 1.0 mV, adult QRS, ST-segment elevation of 0 mV, and a P-R interval of 0.16 seconds
Pacemaker waveform
Paced arrhythmias
- Atrial 80 BPM
- Asynchronous 75 BPM
- Demand with frequent sinus beats
- Demand with occasional sinus beats
- Atrio-ventricular sequential
- Noncapture (one time)
- Nonfunction
Power-on default
Off
Pacer pulse width
0.1, 0.5, 1.0, 1.5, 2.0 ms ± 5%
Pacer pulse
Amplitude
0 (off), 1, 2, 5, 10 mV ± 10 % for lead II (reference lead) with other leads proportional as for performance waves.
Accuracy
Reference lead II: ± (5 % setting + 0.2 mV)
Arrhythmia
Baseline NSR
80 BPM
PVC focus
Left focus, standard timing (except where specified)
Supraventricular arrhythmia
- Atrial fibrillation (coarse or fine)
- atrial flutter sinus arrhythmia
- missed beat (one time)
- atrial tachycardia
- paroxysmal atrial tachycardia
- nodal rhythm
- and supraventricular tachycardia
Premature arrhythmia
- (All one-time events) Premature atrial contraction (PAC)
- premature nodal contraction (PNC)
- PVC1 left ventricular
- PVC1 left ventricular, early
- PVC1 left ventricular, R on T
- PVC2 right ventricularl PVC2 right ventricular, early
- PVC2 right ventricular, R on T
- and multifocal PVCs
Ventricular arrhythmia
- PVCs 6, 12, or 24 per minute
- frequent multifocal PVCs
- bigeminy
- trigeminy
- multiple PVCs (one-time run of 2, 5, or 11 PVCs)
- ventricular tachycardia
- ventricular fibrillation (coarse or fine)
- and asystole
Conduction defect
First-, second-, or third-degree AV block
- and right- or left-bundle branchblock
Power-on default
None (off)
ECG performance testing
Amplitude
0.05 to 0.45 mV (0.05 mV steps), 0.5 to 5.5 mV (0.5 mV steps)
Pulse wave
30, 60 BPM, with 60 ms pulse width
Square wave
2.0, 0.125 Hz
Triangle wave
2.0, 2.5 Hz
Sine wave
0.5, 5, 10, 40, 50, 60, 100 Hz
R-wave detection wave form
Haver-Triangle
R-wave rate
30, 60, 80, 120, 200, and 250 BPM
R-wave width
20 to 200 ms (10 ms steps) Additional Steps: 8, 10, and 12 ms
Rate accuracy
± 1%
Amplitude accuracy
±(2 % of setting + 0.05 mV) , Lead II (Exception: ±5 % for R waves ≤20 ms)
Power-on default
None (off)
Fetal/Maternal ECG (ProSim 3 only)
Fetal heart rate (Fixed)
60, 90, 120, 140, 150, 210 and 240 BPM
Fetal heart rate (IUP)
140 BPM at beginning, then varies with pressure
Intrauterine-pressure waveforms
Early deceleration, late deceleration, and uniform acceleration
Wave duration
90 seconds, bell-shaped pressure curve, from 0 to 90 mmHg and returning to 0
IUP period
2, 3, or 5 minutes; and manual
Power-on default
FHR 120 BPM, early deceleration, manual
Invasive blood pressure
Channels
4, each independently settable with identical parameters and are individually electronically isolated from other signals
Input/output impedance
300 Ω ± 10 %
Exciter input range
2.0 to 16.0 V rms
Exciter-input frequency range
DC to 5000 Hz
Transducer sensitivity
5 or 40 μV/V/mmHg
Pressure accuracy
± 2 % of setting + 2 mmHg (valid for dc excitation only)
Static Levels, Channel 1
-10, 0, 80, 160, 240, 320, 400 mmHg
Static Levels, Channel 2
-10, 0, 50, 100, 150, 200, 240 mmHg
Static Levels, Channel 3 (ProSim 3 only)
-5, 0, 20, 40, 60, 80, 100 mmHg
Static Levels, Channel 4 (ProSim 3 only)
-5, 0, 20, 40, 60, 80, 100 mmHg
Dynamic waveforms, Channel 1
- Arterial: 120/80
- Radial Artery: 120/80
- Left ventricle: 120/00
- Right ventricle: 25/00
Dynamic waveforms, Channel 2
- Arterial: 120/80
- Radial artery: 120/80
- Left ventricle: 120/00
- Right atrium (central venous or CVP): 15/10
- Right ventricle: 25/00
- Pulmonary artery: 25/10
- Pulmonary-artery wedge: 10/2
- Left atrium: 14/4
Dynamic waveforms, Channel 3
- Arterial: 120/80
- Radial artery: 120/80
- Left ventricle: 120/00
- Right atrium (central venous or CVP): 15/10
- Right ventricle: 25/00
- Pulmonary artery: 25/10
- Pulmonary-artery wedge: 10/2
- Left atrium: 14/4
Dynamic waveforms, Channel 4
- Swan-Ganz sequence
- Right atrium (CVP)
- Right ventricle (RV)
- Pulmonary artery (PA)
- Pulmonary-artery wedge (PAW)
Respiration artifact
BP delta changes from 3 to 16 mmHg
Output connector
DIN 5-Pin
Power-on default
0 mmHg
Respiration
Rate
0 (OFF), 15, 20, 30, 40, 60, 80, 100, 120 BrPM
Waves
Normal or ventilated
Ratio (inspiration: expiration)
1:1
Impedance variations (Δ Ω)
0.2, 0.5, 1 or 3 Ω peak-to-peak variation of lead impedance
Accuracy delta
± 10 %
Baseline
500, 1000, 1500, 2000 Ω, Leads I, II, III
Accuracy baseline
± 5 %
Respiration lead
LA or LL
Apnea selection
OFF, 12, 22 or 32 seconds (one-time events), or continuous (Apnea ON = respiration OFF)
Power-on default
20 BrPM, delta 1.0 Ω, 1000 Ω baseline
Temperature
Temperature
0 °C (32 °F), 24 °C (75.2 °F), 37 °F (98.6 °C), and 40 °C (104 °F)
Accuracy
± 0.1° C
Compatibility
Yellow Springs, Inc. (YSI) Series 400 and 700
Output connector
Circular DIN 4-pin
Power-on default
0 °C (42 °F)
Cardiac output
Catheter type
Baxter Edwards, 93a-131-7f
Calibration coefficient
0.542 (0 °C) injectate), 0.595 (24 °C injectate)
Blood temperature
37 °C (98.6 °F) ± 2 %
Injectate volume
10 cc
Injectate temperature
0 °C or 24 °C ± 2 % value
Cardiac output
2.5, 5, 10 liters per minute ± 5 %
Faulty-injectate curve
Waveform for simulation available
Left-to-right shunt curve
Waveform for simulation available
Calibrated pulse
1.5
Output connector
DIN 7-PIN
Power-on default
2.5 liters perminute, 0 °C injectate