Operating:
10 °C to 40 °C (50 °F to 104 °F)
Storage:
-20 °C to +60 °C (-4 °F to +140 °F)
Humidity
10 % to 90 % non-condensing
Altitude
3,000 meters (9,843 ft)
Dimensions (L x W x H)
18 cm x 9.3 cm x 5.5 cm (7.1 in x 3.7 in x 2.2 in)
Display
LCD touch-screen color display
Communication
USB port (for calibration and firmware updates only)
Power
Lithium-ion rechargeable battery
Battery Life
4 hours (minimum), 40 NIBP cycles typical
Safety Standards
IEC 61010-1 3rd Edition; Pollution degree 2
Certifications
CE, CSA, C-TICK N10140 , RoHs
Electromagnetic Compatibility (EMC)
IEC 61326-1:2012
ECG Reference
The ECG amplitudes specified are for Lead II (calibration), from the baseline to the peak of the R wave. All other leads are proportional.
Normal Sinus Rhythm
12-lead configuration with independent outputs referenced to right leg (RL) Output to 10 Universal ECG Jacks, color-coded to AHA and IEC standards.
Amplitude 1.0 mV Other leads are proportional to Lead II (reference lead) in percentage per:
- Lead I: 70
- Lead II: 100
- Lead III: 30
- Lead V1: 24
- Lead V2: 48
- Lead V3: 100
- Lead V4: 120
- Lead V5: 112
- Lead V6: 80
Display
LCD touch-screen color display
Communication
USB port (for calibration and firmware updates only)
Amplitude Accuracy
±5 % of setting Lead II
ECG Rate
30, 60, 80, 90, 120, 150, 180, 210, 240, 270, 300, and 320 BPM (Preset Hypotensive Condition is at 40BPM)
Rate Accuracy
±1 % of setting
ECG Waveform Selection
Adult (80 ms) or neonatal (40 ms) QRS duration
Power-On Default
60 BPM, 1.0 mV, adult QRS
Atrial Fibrillation
Coarse or fine
Premature Ventricular Contraction
Left ventricular
Ventricular Tachycardia
160 or 200 BPM
Ventricular Fibrillation
Coarse or fine
Transvenous Pacer Pulse
75 BPM, left arterial, 3 mV amplitude on lead II, Accuracy ±10 %, 1.0 ms width
2nd Degree AV Block
Type 1
3rd Degree AV Block
3rd degree AV block
Amplitude 1 mV Other leads are proportional to Lead II (reference lead) in percentage per:
- Lead I: 70
- Lead II: 100
- Lead III: 30
- Lead V1: 24
- Lead V2: 48
- Lead V3: 100
- Lead V4: 120
- Lead V5: 112
- Lead V6: 80
Square Wave
60 ms at 2 Hz
Rate
0 (OFF), 10 to 100 BrPM in 10 BrPM steps
Impedance Variations (Δ Ω)
1 Ω
Accuracy Delta
±(10 % + 0.05 ohm)
Baseline
500 Ω to circuit common, giving 1000 Ω between any two leads
Respiration Lead
LA or LL (default)
Channels
1 electrically isolated from all other signals
BP Output
Circular DIN 5-pin
Input/Output Impedance
300 Ω ± 10 %
Exciter Input Range
2 to 16 V peak
Exciter-Input Frequency Range
DC to 5000 Hz
Transducer Sensitivity
5 µV/V/mmHg
Pressure Accuracy
± (1 % of setting + 1 mmHg) Accuracy guaranteed for DC excitation only
Static Pressure
0, 80, 160, and 250 mmHg
Dynamic Waveforms
Synchronization to ECG heartrate
Chambers simulated and systolic/diastolic pressure:
Type
IBP (Arterial) IBP (Left ventrical)
Manometer (Pressure Meter)
- Range 10 to 400 mmHg
- Resolution 0.1 mmHg (for display purposes)
- Accuracy ± (1 % reading + 1 mmHg)
Pressure Source
Inflation bulb or device under test
Pulse
2 mmHg max into 500 ml NIBP system
Volume of air moved
1 ml max
Simulations
Adult: 60/30 (40), 120/80 (93); 150/100 (117); and 200/150 (167)
Neonatal:
35/15 (22) and 70/40 (50)
Repeatability
Within ± 2 mmHg (at maximal pulse size independent of a device under test)
Synchronization
To ECG heartrate (maximal rate 120 BPM)
Target pressure
20 to 400 mmHg
Elapse time
0:30 to 5:00 minutes: seconds in 30 second steps
Leakage rate
1 to 200 mmHg/minute
Pressure Relief Test Range
100 to 400 mmHg
Presets
- Normal
- Hypertensive
- Hypotensive
Autosequences
- Cardiac failure sequence
- Exercise sequence
- Respiration sequence
- Monitor testing sequence