Static Rates:
30, 60, 90, 120, 150, 180, 210,and 240 BPM
ECG Sensitivity:
50 µV, 100 µV, 200 µV, 0.5 mV, 1 mV, and 2 mV
US-1 tracks primary fetal ECG rates. US-2 tracks secondary fetal activity for either independent “normal or “twins simulation, US-2 rate fixed at 140 BPM
Fetal Patterns:
Note: US-1 fetal ECG track these selections. US-2 is in normal pattern, except during TREND #1 selection.
Variability selections (added to fetal ECG)
Absent variability, low variability, mild variability, high variability, severe variability, and long-tern variability
Note:
These patterns repeat and toco channel will perform toco wave selected
Optional mechanical fetal heart
Provides a mechanical interface to the ultrasound transducer; can be connected to either ultrasound channels. This option, due to its power consumption, requires an ac power adapter to be connected.
TACHYCARDIA:
Tachycardia pattern
BRADYCARDIA:
Bradycardia pattern
ARRHYTHMIAS:
Arrhythmia pattern
LATE DECELERATION:
Late deceleration
EARLY DECELERATION:
Early deceleration
MODERATE DECELER:
Moderate variable deceleration
ACCELERATION #1:
Acceleration wave #1
ACCELERATION #2:
Acceleration wave #2
SINUSOIDAL (HIGH):
Sinusoidal pattern, large change
SINUSOIDAL (LOW):
Sinusoidal pattern, small change
SEVERE VAR. DEC. #1:
Severe deceleration wave #2
SEVERE VAR. DEC. #2:
Severe variable deceleration
PROLONGED DECELERATION:
Prolonged deceleration
BIPHASIC DECELERATION:
Biphasic deceleration
EXAGGERATED DECELERATION:
Exaggerated deceleration
NON-UNIFORM DECELERATION:
Non-uniform deceleration
VAR. DECELERATION (U):
Variable deceleration "U" shaped
VAR. DECELERATION TACH:
Variable deceleration with high-rate BPM
VAR. DECELERATION (V):
Variable deceleration "V" shaped
VAR. DECELERATION (POST):
Variable deceleration post exaggerated
VAR. DECELERATION:
Variable deceleration
DECELERATION (POSITION):
Variable deceleration with position changes
LONG DECELERATION:
Long deceleration
COMPENSATORY ACCELATION:
Compensatory acceleration
ECG Static Rates:
60, 80, 100, 120, 140, 160 BPM
ECG Sensitivity:
0.5 mV, 1 mV, and 2 mV
Note:
Pattern selected during Trend #1 selection
Note:
Toco waveform selection not available during Trend #1
EXECUTE WAVEFORM:
Start toco waveform
UTERINE WAVE OFF:
Stop toco waveform
ANALOG 0 TO 1 VOLT:
Analog range 0 to 1 V (1 V = 100 mmHg)
ANALOG 0 TO -1 VOLT:
Analog range 0 to -1 V (1 V = 100 mmHg)
UTERINE WAVE 0 TO 25:
Range of toco waveform
UTERINE WAVE 0 TO 50:
Range of toco waveform
UTERINE WAVE 0 TO 100:
Range of toco waveform
SHORT DURATION:
Toco waveform of short duration
NORMAL DURATION:
Normal duration of toco waveform
INCREASED DURATION:
Long duration of toco waveform
UTERINE LEVEL = ZERO:
Zero toco channel (automatic on power up)
UTERINE STATIC + 20:
Increase toco static level by 20 mmHg (0 to 100 mmHg)
INCR. RESTING TONE:
Resting tone increases
COUPLING:
2 close toco waves
TRIPLING:
3 close toco waves
UTERINE PRESSURE SENSITIVITY:
5 or 40 µV on power up
Important Notes:
US-1 tracks the fetal ECG rates.
US-2 is the second ultrasound channel with anormal fetal ECG pattern.
On the fetal and maternal ECG, the fetal ECGis 1/4 the size of the maternal ECG.
The PS-320 turns on in the following state:
- Fetal ECG static rate @ 150 BPM
- US-1 tracks @ 150 BPM
- US-2 normal pattern
- Pressure sensitivity @ 5 µV/mmHg
- Pressure/Toco set to zero
- Maternal ECG rate @ 80 BPM
- ECG sensitivity @ 1 mV
- Toco wave is normal duration @ 0 to 50 divisions (i.e. 0 to 50 mmHg)
Display:
2-line x 16-character LCD with keypad
RS232:
Bidirectional interface, 9600 Baud
Power:
9V battery/battery eliminator; low battery indication set at 6 V
Dimensions:
6.1 " L x 3.7 " W .x 1.3 " H (15.6 cm L 9.4 cm W x 3.4 cm H)
Temperature:
59 to 95 ˚F (15 to 35 ˚C), operating; 32 to 122 ˚F (0 to 50 ˚C), storage